RegWardenFDA facility compliance records
Recall

Class II — Radiometer Medical ApS

Recalling firmRADIOMETER MEDICAL APS
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe firm has become aware that there is a potential problem relating to the blood gas and immunoassay analyzer Systems that may result in patient mix-up when connected to some third-party devices. The error may lead to serious adverse health consequences for the patient caused by patient data mix-up.
ProductAQURE basic sytem, FLEXLINK, REF 933-599, Version: 2.3.3 (17), CE IVD, Rx Only, UDI: (01)05700699335999 Intended use: The AQURE system is intended to let you manage analytical devices and operator profiles.
Quantity947
DistributionWorldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV; and countries of:  Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, India, Italy, Japan (Tokyo-not a country but it was listed), Netherlands, Norway…
Initiated2019-08-08
Classified2020-01-24
Terminated2022-06-06
Recall numberZ-0864-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Radiometer Medical ApS (Z-0864-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84496/

Not open yet

Alerts for RADIOMETER MEDICAL APS

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.