RegWardenFDA facility compliance records
Recall

Class I — ResMed Ltd.

Recalling firmRESMED CORP.
Registered facilityView this company's 7 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusCompleted
ReasonCombination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of the battery and, that powers on automatically when connected to AC power without pressing the power switch.
ProductResMed Stellar, Non-invasive/invasive ventilators, Model: 100, 150
Quantity69
DistributionU.S.: PA, MA, VA, TN, CT, IL, GA, FL, MT
Initiated2019-12-12
Classified2020-02-06
Recall numberZ-0915-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — ResMed Ltd. (Z-0915-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84551/

Not open yet

Alerts for RESMED CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.