Class II — ulrich medical USA Inc

Recalling firmULRICH MEDICAL USA, INC.
Registered facilityULRICH MEDICAL USA, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonDue to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
ProductUlrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
Quantity179 units
DistributionAZ, TX, GA, FL, AR, CA
Initiated2019-12-31
Classified2020-02-24
Terminated2025-11-17
Recall numberZ-1350-2020
InitiationVoluntary: Firm initiated