RegWardenFDA facility compliance records
Recall

Class II — Trilliant Surgical, LLC

Recalling firmTrilliant Surgical
Registered facilityTrilliant Surgical LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Product8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
Quantity117
DistributionDistributors in 21 states: AR, AZ, CA, FL, GA, IA,IN, KS, KY, MN, MO, NC, ND, NJ, NY, OH, TN, TX, UT, WA, and WI.
Initiated2019-11-04
Classified2020-04-01
Terminated2021-04-28
Recall numberZ-1635-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Trilliant Surgical, LLC (Z-1635-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84987/

Not open yet

Alerts for Trilliant Surgical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.