Class II — LivaNova USA Inc.

Recalling firmLivaNova USA, Inc.
Registered facilityLivaNova USA, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
ProductLivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.
QuantityTotal Worldwide=9309 systems (2713 system in U.S.)
DistributionU.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none
Initiated2020-02-24
Classified2020-03-13
Terminated2022-08-04
Recall numberZ-1498-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LivaNova USA Inc. (Z-1498-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85098/

Not open yet

Alerts for LivaNova USA, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.