RegWardenFDA facility compliance records
Recall

Class II — Hologic, Inc

Recalling firmHOLOGIC, INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonReagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.
ProductPanther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture Reagent - S, Nucleic Acid in a buffered solution and FER-S Fusion Enhancer Reagent-S (Lithium Hydroxide Solution), UDI: 15420045509214
Quantity47 kits
DistributionUS: AR, CA, FL, ID, IL, IN, LA, NM, NY, PA, TX, UT, VI, WA, WI OUS: AU, CA
Initiated2019-10-10
Classified2020-05-28
Recall numberZ-2163-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Hologic, Inc (Z-2163-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85346/