Class II — Encore Medical, LP

Recalling firmENCORE MEDICAL, L.P.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe firm received complaints that the EMPOWR Acetabular Impactor Balls were fracturing/cracking during impaction which could cause debris to migrate or be left in the joint space.
ProductEMPOWR Acetabular System, Non-Sterile, Model Numbers 803-15-023 803-15-026 803-15-027 803-15-028 803-15-029 803-15-030 Product Usage: The EMPOWR Acetabular Impactor Balls is intended for use in joint replacement in patients suffering from disability due to noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; rheumatoid arthritis, correction of functional deformity, femoral fracture, and in the salvage of previously failed surgical attempts.
Quantity86 units
DistributionUS Nationwide distribution including in the states of MN, VA, IL, IN, SC, N, TX, CA.
Initiated2019-10-22
Classified2020-05-08
Recall numberZ-1929-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Encore Medical, LP (Z-1929-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85376/

Not open yet

Alerts for ENCORE MEDICAL, L.P.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.