Class II — Radiometer Medical ApS
| Recalling firm | RADIOMETER MEDICAL APS |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors. |
| Product | ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. |
| Quantity | 1233 units |
| Distribution | Worldwide distribution - US Nationwide. Also distributed OUS. |
| Initiated | 2019-11-21 |
| Classified | 2020-05-28 |
| Terminated | 2024-11-08 |
| Recall number | Z-2139-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Radiometer Medical ApS (Z-2139-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85390/