RegWardenFDA facility compliance records
Recall

Class II — Beckman Coulter Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.
ProductiQ200 Series Urine Microscopy Analyzer- All part numbers
Quantity3576 units
DistributionUS Distribution: Nationwide Foreign Distribution: Worldwide
Initiated2020-03-30
Classified2020-05-08
Terminated2023-08-31
Recall numberZ-1945-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter Inc. (Z-1945-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85423/