RegWardenFDA facility compliance records
Recall

Class II — Uromedica Inc.

Recalling firmUROMEDICA, INC.
Registered facilityUROMEDICA, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonPotential issue with the manufacture and inspection of U-channel sheaths in Implantation Instruments Sets. Affected U-channel sheaths may be more likely to cause damage to ProACT implant devices when passes through the sheaths.
ProductProACT Implantation Instrument Set, is a reusable, steam sterilizable surgical instrument set consisting of a U-Channel Sheath through which the Pro ACT balloon implant is introduced.
Quantity11 tool sets
DistributionNationwide distribution to the following states: CA, GA, FL, MI, and VA.
Initiated2019-04-10
Classified2020-04-28
Terminated2020-05-18
Recall numberZ-1808-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Uromedica Inc. (Z-1808-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85486/