Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
ProductCios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194
Quantity20
DistributionUS Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.
Initiated2020-04-17
Classified2020-05-26
Terminated2020-08-25
Recall numberZ-2128-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-2128-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85572/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.