RegWardenFDA facility compliance records
Recall

Class II — Defibtech, LLC

Recalling firmDEFIBTECH, LLC
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA component issue may cause the AED to abort a shock delivery or reset unexpectedly, which may cause an affected device to fail to delivery shock, and/or to fail to deliver shock in range.
ProductDDU-2400EN Lifeline PRO, UDI 00815098020270 - Product Usage: Automated External Defibrillators (AED) are indicated for use on victims of sudden cardiac arrest (SCA) who are: - unconscious and unresponsive, not breathing or not breathing normally.
Quantity860
DistributionWorldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Malaysia, Mauritius, Netherlands, Thailand, United Arab Emirates.
Initiated2020-04-27
Classified2020-06-17
Terminated2021-01-08
Recall numberZ-2393-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Defibtech, LLC (Z-2393-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85702/