Class II — RAYSEARCH LABORATORIES AB
| Recalling firm | RaySearch Laboratories AB (publ) |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat. |
| Product | RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2 |
| Quantity | 12 |
| Distribution | Distributed to one consignee in Nashville. Foreign distribution to one consignee in Belgium. |
| Initiated | 2020-05-26 |
| Classified | 2020-06-24 |
| Recall number | Z-2426-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — RAYSEARCH LABORATORIES AB (Z-2426-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85796/
Alerts for RaySearch Laboratories AB (publ)
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.