Recall

Class II — Steris Corporation

Recalling firmSTERIS Corporation
Registered facilityKey Surgical, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonA recent FDA inspection at the firm identified that the foil test performed during installation of the device does not have clear acceptance criteria.
ProductInnoWave Sonic Irrigator, There are currently six models (PCF, Pro, Ultra, Ultra+, Unity, and ECO) - Product Usage: STERIS InnoWave Sonic Irrigators are intended to thoroughly clean (remove tissue, blood, and other contaminants from) a variety of reusable surgical instruments. The equipment provides sonic irrigation (ultrasonic cavitation) to both interior and exterior of lumened instruments.
Quantity1283
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, and United Arab Emirates.
Initiated2020-06-26
Classified2020-07-24
Terminated2020-12-22
Recall numberZ-2700-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Steris Corporation (Z-2700-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/85871/