Class II — Sutter Medizintechnik GmbH
| Recalling firm | Sutter Medizintechnik GmbH |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The label on the cardbox and the blister show the correct symbols for sterilization using irradiation and storage at 5 C to 40 C, the package insert (IFU) is labelled incorrectly. It shows the symbol for sterilization using Ethylene Oxide with the description for sterilization using irradiation. |
| Product | Sutter Swyng non-stick bipolar forceps - Product Usage: single-use re electrosurgical instruments. The bipolar for-ceps are provided with bayonet-style handle design with straight or angled tips and different total lengths. They are to be connected through an appropriate bipolar cable with the bipolar output of an electrosurgical generator. The electrodes are provided sterile and are single-use instruments. |
| Quantity | 1337 units |
| Distribution | US Nationwide distribution including in the states of MD, NJ, MI, NC, IN, CA, AL, WA, OK, MS, GA, FL, SC, TX, CO, MN, MO, KS. |
| Initiated | 2020-06-23 |
| Classified | 2020-07-31 |
| Terminated | 2020-08-21 |
| Recall number | Z-2744-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Sutter Medizintechnik GmbH (Z-2744-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86035/
Alerts for Sutter Medizintechnik GmbH
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.