RegWardenFDA facility compliance records
Recall

Class II — Clinical Innovations, LLC

Recalling firmClinical Innovations, LLC
Registered facilityCLINICAL INNOVATIONS, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonPossible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.
ProductClinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025
Quantity19450 units/kits
DistributionUS: CO, IL, IN, MO, OH, RI, UT, and VA. OUS: None
Initiated2020-08-13
Classified2020-12-07
Terminated2024-05-24
Recall numberZ-0541-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Clinical Innovations, LLC (Z-0541-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86255/

Not open yet

Alerts for Clinical Innovations, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.