Class III — Roche Diagnostics Operations, Inc.
| Recalling firm | Roche Diagnostics Operations Inc. |
|---|---|
| Registered facility | View this company's 14 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class III |
| Status | Terminated |
| Reason | Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents. |
| Product | Albumin Gen.2, Catalog 05166861190 |
| Quantity | 11963 |
| Distribution | Domestic distribution nationwide. |
| Initiated | 2020-08-27 |
| Classified | 2020-10-13 |
| Terminated | 2024-05-22 |
| Recall number | Z-0104-2021 |
| Initiation | Voluntary: Firm initiated |