RegWardenFDA facility compliance records
Recall

Class II — Maquet Cardiovascular Us Sales, Llc

Recalling firmMaquet Cardiovascular LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPlastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
ProductYuno II Mobile Operating Table - Product Usage: support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143302F0 (US) Part 143302B0 (OUS)
Quantity6 units US
DistributionWorldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, MACAU, MACEDONIA, MALAYSIA, MEXICO, NEW ZEALAND, OMAN, POLAND, QATAR, RO…
Initiated2020-09-24
Classified2020-10-20
Terminated2023-02-28
Recall numberZ-0133-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Maquet Cardiovascular Us Sales, Llc (Z-0133-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86495/

Not open yet

Alerts for Maquet Cardiovascular LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.