Recall

Class II — Westmed, Inc.

Recalling firmWESTMED, INC.
Registered facilityWESTMED, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
ProductREF 0562K (Adult Elongated O2/CO2 Mask - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
Quantity16,000 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
Initiated2020-11-11
Classified2020-12-29
Terminated2023-02-07
Recall numberZ-0745-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Westmed, Inc. (Z-0745-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/86903/

Not open yet

Alerts for WESTMED, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.