RegWardenFDA facility compliance records
Recall

Class II — Bioventus, LLC

Recalling firmBioventus LLC
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonProduct may contain white particulates that are microbial in nature.
ProductEXOGEN Ultrasound Coupling Gel
Quantity115,584 units
DistributionUnited States (Nationwide), Australia, Austria, Belgium, Canada, Cyprus, Denmark, France, Germany, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, United Kingdom
Initiated2020-12-17
Classified2021-01-09
Terminated2024-08-07
Recall numberZ-0810-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Bioventus, LLC (Z-0810-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87002/

Not open yet

Alerts for Bioventus LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.