Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits. |
| Product | Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno |
| Quantity | 5602 |
| Distribution | Worldwide-US Nationwide |
| Initiated | 2020-11-30 |
| Classified | 2021-01-11 |
| Recall number | Z-0656-2021 |
| Initiation | FDA Mandated |