RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPotential heat and humidity exposure. The affected lot contained an atypical sandy, flaky or grainy consistency when mixing.
ProductVERSABOND AB 40 GRAMS FORMULATION 2, antibiotic infused bone cement, Product Number: 71271440 - Product Usage: VERSABOND is a medium viscosity bone cement which can be used for hip, knee, and shoulder implant fixation. VERSABOND AB includes the antibiotic- gentamycin.
Quantity5190 devices
DistributionWorldwide distribution - US Nationwide distribution in the state of OH and the countries of Dubai, Canada, Korea, Chile, Puerto Rico, Panama, and Costa Rica.
Initiated2020-12-18
Classified2021-01-29
Terminated2023-05-01
Recall numberZ-0933-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew, Inc. (Z-0933-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87035/