Class II — Qiagen Sciences LLC

Recalling firmQIAGEN Sciences LLC
Registered facilityQIAGEN (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonKit may generate false Q546R mutation positive results caused by nonspecific molecular interactions within the Q546R higher than previously observed and described within the Instructions For Use.
Producttherascreen PIK3CA RGQ PCR Kit (23) US IVD product- aid clinicians in identifying breast cancer patients who may be eligible for treatment with PIQRAY (alpelisib) based on a PIK3CA Mutation Detected result. REF 873121
Quantity575 kits US
DistributionCA, IN, MI, MN, NC, NJ, NM, OH, TX
Initiated2020-12-11
Classified2021-01-28
Terminated2022-04-22
Recall numberZ-0937-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Qiagen Sciences LLC (Z-0937-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87047/