Class II — Roche Diagnostics Operations, Inc.

Recalling firmRoche Diagnostics Operations Inc.
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonMay display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.
Product9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.
Quantity1,214 units
DistributionUS Nationwide distribution.
Initiated2021-01-06
Classified2021-02-08
Terminated2024-02-28
Recall numberZ-1098-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-1098-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87102/

Not open yet

Alerts for Roche Diagnostics Operations Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.