Class II — Becton Dickinson & Co.
| Recalling firm | Becton Dickinson and Company |
|---|---|
| Registered facility | View this company's 36 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined. |
| Product | BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider. |
| Quantity | 938 kits |
| Distribution | US nationwide distribution. |
| Initiated | 2021-01-07 |
| Classified | 2021-02-05 |
| Terminated | 2021-12-03 |
| Recall number | Z-0954-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Becton Dickinson & Co. (Z-0954-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87178/
Alerts for Becton Dickinson and Company
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.