RegWardenFDA facility compliance records
Recall

Class II — Stryker Communications

Recalling firmSTRYKER CORP.
Registered facilityView this company's 61 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThe manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.
ProductS-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD
Quantity1,393 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
Initiated2021-01-29
Classified2021-03-24
Recall numberZ-1297-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stryker Communications (Z-1297-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87331/

Not open yet

Alerts for STRYKER CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.