RegWardenFDA facility compliance records
Recall

Class II — Beckman Coulter Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonInstruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
ProductMicroScan WalkAway-96 plus Reconditioned Instrument
Quantity9 systems
DistributionU.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.
Initiated2019-09-04
Classified2021-04-07
Terminated2023-01-11
Recall numberZ-1373-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter Inc. (Z-1373-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87486/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.