Class II — Siemens Medical Solutions USA, Inc
| Recalling firm | SIEMENS MEDICAL SOLUTIONS USA, INC. |
|---|---|
| Registered facility | View this company's 9 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited functionality of the Artis system up to system failure. Unintended movement of the C-Arm may cause a hazardous situation to the patient, operator, or staff members and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system |
| Product | Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Model Numbers: System Material # AXIOM Artis FC 05904433; AXIOM Artis FA 05904441; AXIOM Artis MP 05904466; AXIOM Artis BC 05904649; AXIOM Artis BA 05904656; AXIOM Artis dBC Mag. Nav. 05917054; AXIOM Artis TA 07007755; AXIOM Artis dTA 07008605; AXIOM Artis dFC 07412807; AXIOM Artis dTC 07413078; AXIOM Artis dBA 07555357; AXIOM Artis dMP 07555365; AXIOM Artis dFA 0755537… |
| Quantity | 3,320 units U.S. |
| Distribution | US Nationwide distribution. |
| Initiated | 2021-03-30 |
| Classified | 2021-04-26 |
| Terminated | 2022-07-12 |
| Recall number | Z-1496-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1496-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/87683/
Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.