Recall

Class I — Baxter Healthcare Corporation

Recalling firmBAXTER HEALTHCARE CORPORATION
Registered facilityView this company's 32 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass I
StatusTerminated
ReasonSystem errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
ProductSpectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Quantity130135 units
DistributionUnited States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Initiated2021-06-04
Classified2021-08-06
Terminated2024-09-25
Recall numberZ-2110-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class I recall — Baxter Healthcare Corporation (Z-2110-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88073/

Not open yet

Alerts for BAXTER HEALTHCARE CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.