Class I — Boston Scientific Corporation
| Recalling firm | Boston Scientific Corporation |
|---|---|
| Registered facility | View this company's 37 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population |
| Product | ADVANTIO DR EL (pacemaker), Model Numbers: a) J064 b) J067 c) K064 (added 7/15/2021) d) K084 (added 7/15/2021) e) K087 (added 7/15/2021) |
| Quantity | 15,717 units |
| Distribution | Worldwide distribution. |
| Initiated | 2021-06-03 |
| Classified | 2021-07-12 |
| Recall number | Z-2015-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Boston Scientific Corporation (Z-2015-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88107/
Alerts for Boston Scientific Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.