Class II — Abbott Laboratories
| Recalling firm | ABBOTT LABORATORIES |
|---|---|
| Registered facility | View this company's 64 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results. |
| Product | REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569 |
| Quantity | 4,727 units |
| Distribution | Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG… |
| Initiated | 2021-06-10 |
| Classified | 2021-08-02 |
| Recall number | Z-2184-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Abbott Laboratories (Z-2184-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88123/
Alerts for ABBOTT LABORATORIES
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?