Class I — Baxter Healthcare Corporation
| Recalling firm | BAXTER HEALTHCARE CORPORATION |
|---|---|
| Registered facility | View this company's 32 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Terminated |
| Reason | Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump. |
| Product | Dose IQ Safety Software used with Spectrum IQ Infusion Pump |
| Quantity | 61 units |
| Distribution | US Nationwide Distribution. |
| Initiated | 2021-07-07 |
| Classified | 2021-08-11 |
| Terminated | 2024-06-20 |
| Recall number | Z-2165-2021 |
| Initiation | Voluntary: Firm initiated |