Class II — CooperSurgical, Inc.
| Recalling firm | COOPERSURGICAL, INC. |
|---|---|
| Registered facility | View this company's 6 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage. |
| Product | Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use in the following incubator and work chamber products; G185 Standard 110V, G185 SensorTech 110V, G185 Standard 230V, G185 SensorTech 230V, G210 InviCell Standard, G210 InviCell Standard US, G210 Invicell Plus Optional pH and External Monitoring, G603 Work Chamber 115V, & G603 Work Chamber 230V. [CooperSurgical part numbers K22400-110, K22200-110, K22400-230, K22200-230, K59500, K59700, K60000, 3.641.831.153, & 3.641.832.153]. |
| Quantity | 421 units |
| Distribution | Only Foreign: Bahrain, Belgium, Bulgaria, Czech Republic Denmark, Egypt, Estonia, France, Georgia, Germany, Ghana, Great Britain, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Netherlands, Norway, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab E… |
| Initiated | 2021-06-30 |
| Classified | 2021-08-16 |
| Terminated | 2022-11-21 |
| Recall number | Z-2307-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — CooperSurgical, Inc. (Z-2307-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88277/
Alerts for COOPERSURGICAL, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.