Class I — Cordis Corporation
| Recalling firm | Cordis US Corp. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Terminated |
| Reason | Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement. |
| Product | Cordis 5F SUPER TORQUE PIG PIGTAIL SPECIAL, Angiographic Catheter, REF SRD7040MB |
| Quantity | 1610 units |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe. |
| Initiated | 2021-07-21 |
| Classified | 2021-09-15 |
| Terminated | 2024-09-09 |
| Recall number | Z-2429-2021 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Cordis Corporation (Z-2429-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88372/