RegWardenFDA facility compliance records
Recall

Class II — MY01, INC.

Recalling firmMY01, Inc.
Registered facilityMY01, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is an incorrect version of the needle in the introducer part of devices. The affected Lots were manufactured using needles with a needle trocar geometry which is out of specification and may require a slightly higher push force from the user to penetrate the patient skin.
ProductMY01 Continuous Compartmental Pressure Monitor, Reference Number MY01-0001
Quantity37 units
DistributionUS Nationwide distribution in the states of MD, OH, CA, NY.
Initiated2021-06-28
Classified2021-08-27
Terminated2022-07-25
Recall numberZ-2384-2021
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — MY01, INC. (Z-2384-2021).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88396/