RegWardenFDA facility compliance records
Recall

Class II — Fort Defiance Industries, LLC

Recalling firmFORT DEFIANCE INDUSTRIES
Registered facilityFORT DEFIANCE INDUSTRIES LLC (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThe rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
ProductP2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Quantity380 devices
DistributionDistributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
Initiated2021-08-19
Classified2021-10-15
Recall numberZ-0121-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Fort Defiance Industries, LLC (Z-0121-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88628/