RegWardenFDA facility compliance records
Recall

Class II — Medicalplastic S.R.L.

Recalling firmMEDICALPLASTIC
Registered facilityMEDICALPLASTIC (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonMedicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
ProductIlizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Quantity581 units
DistributionMemphis, TN
Initiated2021-06-17
Classified2021-11-17
Recall numberZ-0245-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Medicalplastic S.R.L. (Z-0245-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88785/

Not open yet

Alerts for MEDICALPLASTIC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.