Recall

Class II — FujiFilm Healthcare Americas Corporation

Recalling firmFUJIFILM Healthcare Americas Corporation
Registered facilityFUJIFILM Healthcare Americas Corporation (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonWhen the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual
ProductArietta 65 Ultrasound Software Versions: V1.0 through V4.0.0
Quantity48 US
DistributionUS Nationwide distribution.
Initiated2021-10-14
Classified2021-11-02
Recall numberZ-0192-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — FujiFilm Healthcare Americas Corporation (Z-0192-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88849/

Not open yet

Alerts for FUJIFILM Healthcare Americas Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.