Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonSystem operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs
ProductSensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies
Quantity403 units
DistributionNationwide
Initiated2021-10-05
Classified2021-11-16
Recall numberZ-0226-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-0226-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/88870/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.