RegWardenFDA facility compliance records
Recall

Class II — Flower Orthopedics Corporation

Recalling firmFLOWER ORTHOPEDICS CORPORATION
Registered facilityFLOWER ORTHOPEDICS CORPORATION DBA Conventus Flower (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonMating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available
ProductFlex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System- Catalog Number: 8435-1
Quantity48 units
DistributionUS Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA, MD, NC, TX.
Initiated2021-11-02
Classified2022-01-03
Terminated2023-10-12
Recall numberZ-0448-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Flower Orthopedics Corporation (Z-0448-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89092/

Not open yet

Alerts for FLOWER ORTHOPEDICS CORPORATION

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.