| Recalling firm | MedShape, Incorporated |
|---|---|
| Registered facility | MEDSHAPE, INC. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | The firm identified a nonconformance in Lot 04882 involving the endcap being manufactured to a longer length than indicated in product specifications, introducing the risk of the +6 mm endcap not seating flush with the nail while inserting during a procedure. |
| Product | DYNANAIL MINI, Endcap +6mm, REF 2600-05-006, Rx Only, Sterile |
| Quantity | 26 units (1 count) |
| Distribution | Domestic: CA, FL, GA, IL, IN, MA, MD, TX, and WA. |
| Initiated | 2021-11-17 |
| Classified | 2022-01-04 |
| Recall number | Z-0453-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)