RegWardenFDA facility compliance records
Recall

Class II — Qiagen Sciences LLC

Recalling firmQIAGEN Sciences LLC
Registered facilityQIAGEN (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.
ProductQIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Quantity376 kits
DistributionDomestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.
Initiated2021-11-08
Classified2022-01-05
Terminated2023-03-29
Recall numberZ-0455-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Qiagen Sciences LLC (Z-0455-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89130/