Class II — Draeger Medical, Inc.

Recalling firmDraeger Medical Systems, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
Reason26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
ProductDraeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Quantity5
DistributionUS Nationwide distribution in the states of NY, PA, and WI.
Initiated2021-12-16
Classified2022-01-14
Terminated2023-04-17
Recall numberZ-0501-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Draeger Medical, Inc. (Z-0501-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89243/

Not open yet

Alerts for Draeger Medical Systems, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.