RegWardenFDA facility compliance records
Recall

Class II — Luminex Corporation

Recalling firmLuminex Corporation
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonImpacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
ProductARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Quantity329
DistributionWorldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.
Initiated2021-11-17
Classified2022-02-02
Terminated2024-07-08
Recall numberZ-0553-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Luminex Corporation (Z-0553-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89244/

Not open yet

Alerts for Luminex Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.