Class II — Compass Health Brands (Corporate Office)
| Recalling firm | Compass Health Brands (Corporate Office) |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | Wheel spoke may crack causing the wheel to separate from the axle may result in patient injuries, including bruising and lacerations requiring stitches. |
| Product | Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) RLEU10BL Voyager Rollator Cobalt Blue UPC: 815067078066; RLEU10PK Voyager Rollator Rose Gold UPC: 815067078073; RLEU10WT Voyager Rollator Ice Palace UPC: 815067078080 |
| Quantity | 3734 units |
| Distribution | Nationwide |
| Initiated | 2022-01-10 |
| Classified | 2022-01-19 |
| Recall number | Z-0510-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Compass Health Brands (Corporate Office) (Z-0510-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89389/
Alerts for Compass Health Brands (Corporate Office)
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.