Class I — Atrium Medical Corporation
| Recalling firm | ATRIUM MEDICAL CORPORATION |
|---|---|
| Registered facility | Atrium Medical Corporation (the firm's only registered facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and contrast which may impact patients' renal function. Potential harms include occlusion or embolism and associated response, with specific outcomes such as amputation, embolism, loss of organ function, organ infarction, or tissue infarction. ***Updated June 2023*** Additional surgical stress caused by prolonged interventional surgery has the potential to lead to myocardial infarction or death. |
| Product | iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420 85424 85440 85441 85442 85443 85444 85445 85450 85451 85452 85453 85454 85455 Product Code/UDI /Description: 85400 00650862854008 iCAST COVERED STENT,5MMX38MMX80CM 85401 00650862854015 iCAST COVERED STENT,5MMX59MMX80CM 85402 00650862854022 iCAST COVERED STENT,6MMX38MMX80CM 85403 00650862854039 iCAST COVERED STENT,6MMX59MMX80CM 85404 00650862854046 iCAST COVERED STENT,7MMX… |
| Quantity | 68812 units |
| Distribution | Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Ireland, Republic of Israel Italy Japan Jordan Kuwait Latvia Lebanon Lithuania Macedonia Malaysia Malta Mexico Neth… |
| Initiated | 2022-03-03 |
| Classified | 2022-05-18 |
| Recall number | Z-1077-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Atrium Medical Corporation (Z-1077-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89852/
Alerts for ATRIUM MEDICAL CORPORATION
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.