Recall
Class II — Abbott
| Recalling firm | ABBOTT LABORATORIES |
|---|---|
| Registered facility | View this company's 64 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device. |
| Product | Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile. |
| Quantity | 588 devices |
| Distribution | Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom. |
| Initiated | 2022-03-24 |
| Classified | 2022-04-19 |
| Terminated | 2026-08-04 |
| Recall number | Z-0957-2022 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Abbott (Z-0957-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89872/
Not open yet