RegWardenFDA facility compliance records
Recall

Class II — Accriva Diagnostics, Inc.

Recalling firmAccriva Diagnostics, Inc.
Registered facilityAccriva Diagnostics, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonTwo whole blood Platelet Reactivity tests exist that share the same reagent formulation and are used to obtain P2Y12 Reactivity Unit results. The test without US-FDA market clearance was distributed and it displays BASE results not displayed by the US-FDA market cleared device. A formula is being provided to health care personnel to calculate the percentage of platelet aggregation inhibition.
ProductThe VerifyNow P2Y12 Test is designed to measure platelet P2Y12 receptor blockade. Substances known to specifically block the P2Y12 receptor include ticagrelor and the thienopyridine class of drugs, including clopidogrel and prasugrel.1-8 The test is based upon the ability of activated platelets to bind fibrinogen. Fibrinogen-coated microparticles aggregate in whole blood in proportion to the number of expressed platelet GP IIb/IIIa receptors. Light transmittance increases as activated platelets bind and aggregate fibrinogen-coated beads. The instrument measures this change in optical signal an…
Quantity11
DistributionUS Nationwide distribution in the state of California.
Initiated2022-03-30
Classified2022-05-20
Terminated2024-11-07
Recall numberZ-1121-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Accriva Diagnostics, Inc. (Z-1121-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/89936/

Not open yet

Alerts for Accriva Diagnostics, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.