RegWardenFDA facility compliance records
Recall

Class II — Dynex Technologies, Inc.

Recalling firmDynex Technologies, Inc.
Registered facilityDYNEX TECHNOLOGIES, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonControl samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.
ProductDYNEX Agility, Agility Analyzer, Model No. 67000
Quantity218 units (17 installed with v1.4.3)
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, CA, DC, FL, GA, IL, IN, MA, MD, MI, NE, NJ, NY, OH, PA, PR, TN, TX, UT, VA, WI, WV and the countries of Belgium, Brazil, China, Czech Republic, France, Germany, Mexico, Qatar, Romania, Russia, South Korea, Switzerland, Taiwan, and UK.
Initiated2020-07-05
Classified2022-06-02
Recall numberZ-1222-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Dynex Technologies, Inc. (Z-1222-2022).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/90099/

Not open yet

Alerts for Dynex Technologies, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.