Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
Reasonif the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm
ProductArtis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700
Quantity13 units
DistributionUS Nationwide distribution.
Initiated2022-04-11
Classified2022-06-07
Recall numberZ-1246-2022
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)